SPECIFICATIONS
Primary Gases
(Performance
approved gases
)
1, 2
Optional Gases
Range
Recommended
Alarm Setting
Recommended
Minimum Alarm
Setting
Path Length
Speed of Response
Analog Output Signal
Environmental
Operating
Temperature
Storage Temperature
Range
Power Supply
Power Consumption
(average RMS)
* Searchline Excel Plus & Edge will respond to most hydrocarbon gases as most do absorb IR in the required wavelengths.
Searchline Excel Plus & Edge will respond to a wide range of Hydrocarbon Gases*:
Methane
, Ethane
, Propane
1,2
1
FM Performance Approved;
1
Calibration to other gases is optional and should be consulted with Honeywell.
Methane
0–5 LEL.m / 0–220,000 ppm.m
Ethane
0–5 LEL.m / 0–120,000 ppm.m
Propane
0–5 LEL.m / 0–85,000 ppm.m
Butane
0–5 LEL.m / 0–70,000 ppm.m
Pentane
0–5 LEL.m / 0–55,000 ppm.m
Hexane
0–5 LEL.m / 0–50,000 ppm.m
Ethylene
0–5 LEL.m / 0–115,000 ppm.m
Propylene
0–5 LEL.m / 0–100,000 ppm.m
Linearized Low 1 LEL.m:
(and equivalent ppm.m)
Methane
44,000 ppm.m
Ethane
24,000 ppm.m
Propane
17,000 ppm.m
Butane
14,000 ppm.m
Pentane
11,000 ppm.m
Hexane
10,000 ppm.m
Ethylene
23,000 ppm.m
Propylene
20,000 ppm.m
Linearized 0 to 0.5 LELm:
(and equivalent ppm.m)
Methane
22,000 ppm.m
Ethane
12,000 ppm.m
Propane
8,500 ppm.m
Butane
7,000 ppm.m
Pentane
5,500 ppm.m
Hexane
5,000 ppm.m
Ethylene
11,500 ppm.m
Propylene
10,000 ppm.m
Searchline Excel Plus: 2 m to 120 m (6 ft to 393 ft)
T90 less than 3 seconds
4–20 mA Normal operation, 1 mA Fault*,
2 mA Inhibit, 3 mA Warning, 21 mA Over-range,
*Searchline Excel Plus will indicate Fault after 30 s without signal. This value is
programmable. To comply with FM performance approval, the maximum value for
Time to Block is 200 s.
-55 °C to +75 °C (-67 °F to +167 °F); DNVGL-CG-0339 Location Class B
-55 °C to +75 °C (-67 °F to +167 °F); DNVGL-CG-0339 Location Class B
24 Vdc nominal (18 Vdc to 32 Vdc)
Transmitter, for ambient operating temperature:
-55°C (-67°F) to -30°C (-22°F)
-30°C (-22°F) to +30°C (+86°F)
above +30°C (+86°F)
Receiver, for ambient operating temperature:
-55°C (-67°F) to -30°C (-22°F)
-30°C (-22°F) to +30°C (+86°F)
above +30°C (+86°F)
, Butane
, Pentane
, Hexane, Ethylene
1,2
1
1
DNV approved
2
Linearized High 3 LEL.m:
(and equivalent ppm.m)
Methane
132,000 ppm.m
Ethane
72,000 ppm.m
Propane
51,000 ppm.m
Butane
42,000 ppm.m
Pentane
33,000 ppm.m
Hexane
30,000 ppm.m
Ethylene
69,000 ppm.m
Propylene
60,000 ppm.m
11.8 W
5.6 W
3.8 W
7.4 W
4.8 W
3.0 W
(9)
, Propylene
,
1
1