Steam sterilization with fractionated pre-vacuum
Pre-vacuum cycles
Medium
Holding time
Sterilization temperature
Drying time
5.7
Storage and Transport
Products
• must be stored in a clean, cool and dry location,
• must be protected against mechanical damage,
• must be handled with great care. Do not drop or throw.
Approved sterilization packaging (e.g. conforming to EN 868, ISO 11607) must be used for sterilization,
subsequent transportation and storage.
In case of return shipments, send only clean and disinfected products in sterile packaging.
6 Disposal
6.1
Packaging
Gebrüder Martin will take back the full packaging upon request. Whenever possible, parts of the packaging
will be reused.
If you do not wish to make use of this offer, you can dispose of the packaging as paper and household
waste.
6.2
Product
After expiry of the product lifetime/explantation, the products must be correctly disposed of or fed into a
recycling system.
6.3
National Regulations
The national regulations and disposal provisions must be observed for all disposal measures.
38
Procedure 1
3 ×
Saturated water vapor
5 min
134 °C, temperature band up to
137 °C
20 min
Instructions for Use
L1® storage system
Procedure 2
3 ×
Saturated water vapor
3 min
132 °C, temperature band up to
135 °C
20 min
Revision 2