Physical and chemical measurements
Spectrophotometer USB / Bluetooth
Ref. 705414 – 705432
3.2.2 Absorbance tab > Kinetic
For the study of slow reactions causing a
change in colored solution, such as the
oxidation
spectrophotometer
determine the kinetics of this reaction.
To do this, select Kinetics. the window
shown opposite appears.
The kinetics can be studied at two different
wavelengths. If the study is at a single
wavelength - a single value is expected.
Select the time interval between two
acquisitions, then start the acquisition.
You can then stop acquisition by clicking on
Stop.
In the case of automatic stop, you can pre-
select the acquisition duration. It will
automatically stop after the set time (in
seconds).
3.3 Fluorescence tab
ENGLISH
of
iodide
ions,
can
be
used
21
the
to
The
fluorescence
fluorescence
emitted
following
its
excitation
wavelength.
Two wavelength ranges are available. It is
also possible to emit no excitation and still
collect
the
spontaneous
emission from the sample.
Before any fluorescence measurement, you
need to perform a calibration. To do this,
insert a cuvette filled with solvent and click on
Calibrate, the
instrument acquires the
reference spectrum. Blackness measurement
follows this calibration. This step determines
the level of noise (optical, electronic...) to be
subtracted
from
measurement.
mode
measures
the
by
your
sample
by
a
selected
fluorescence
the
curve
during