5.3 Balance Setup

Enter this sub-menu to customize the balance functionality.
Note: The factory default settings are shown below in bold.

5.3.1 Filter Level

Set the amount of signal filtering.
Low
= faster stabilization time with less stability.
Medium = normal stabilization time with normal stability.
High
= slower stabilization time with more stability.
LowW
Low

5.3.2 AZT (Auto Zero Tracking)

Set the automatic zero tracking functionality.
Off = disabled.
0.5d = display maintains zero up to a drift of 0.5 graduation per second.
1d
= display maintains zero up to a drift of 1 graduation per second.
3d
= display maintains zero up to a drift of 3 graduations per second.
0.5 d
0.5 d

5.3.3 Auto Tare

Set the automatic tare.
When Automatic Tare is set to On, the first object placed on the pan will be deemed as a container and tared
automatically.
Off = disabled.
On = enabled.
A.tArE
Auto Tare
mMed
Medium
1 d
1 d
OFF
Off
FiltEr
HIGH
High
AZt
Auto Zero Tracking
3 d
3 d
ON
On
EN-17