Télécharger Imprimer la page

Agilent Technologies Turbo-V 551 Rack Notice De Mode D'emploi page 241

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
N.
167
200
201
202
203
204
205
206
211
222
232
300
301
302
Turbo-V 551 Rack Controller User Manual / 87-901-031-01
Read/
Data
Description
Write
Type
R/W
L
Stop Speed Reading (when
active, the pump rotation
speed is measured also in
Braking status)
R
N
Pump
current expressed in mA dc
R
N
3 phase voltage expressed in
V
rms
R
N
Output power expressed in W
R
N
Driving frequency expressed
in Hz
R
N
Pump bearing temperature
expressed
in °C
R
N
System status
R
N
Error code
R
N
Controller Internal temperature
in °C
R
N
Body pump temperature in °C
R
N
Rotational frequency in Hz
R
N
Cycle time in minutes (zeroed
by the reset command)
R
N
Cycle number (zeroed by the
reset command)
R
N
Pump life in hours (zeroed by
the reset command)
Technical Information
Window Meanings
Admitted Values
0 = No
1 = Yes
(default = 1)
Stop = 0
Waiting intlk = 1
Ramp-Up = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see the
following figure
0 to 999999
0 to 999999
0 to 999999
(A)
17
241/288