Technical Information
Window Meanings
N.
126
143
144
145
147
155
157
240/288
Read/
Data
Description
Write
Type
R/W
N
Vent valve opening delay
expressed in 0.2 sec
R/W
N
External Fan Configuration
R/W
L
External Fan activation
R/W
L
Purge Valve open/close (N.C.)
R/W
N
Vent Valve opening (Time
expressed in 0.2 s)
R
N
Power Limit Applied expressed
in W
R/W
L
Gas Load Type
Turbo-V 551 Rack Controller User Manual / 87-901-031-01
Admitted Values
0 to 65535
Corresponding to 0 to
13107 sec
(Default = 15)
0 = Always ON
1 = Automatic
2 = Serial
(default = 2)
0 = OFF
1 = ON
(default = 0)
0 = close
1 = open
(Default =1)
0 to 65535 bit
0 = infinite
min 0.2 s,
Max 13107 s
(default = 0)
0 = Ar
1 = N
2
(default = 0)