Turbo-V 551 Rack Controller User Manual / 87-901-031-01
Read/
Data
Description
Write
Type
R/W
L
Cooling agent
R/W
L
Active Stop
(can be written only in Stop
condition)
R/W
N
Baud rate
R/W
L
Interlock mode (if configured
as Impluse the Interlock signal
acts only at the start)
R/W
L
Analog output link: the output
voltage can be related to
different quantities.
R/W
N
Low speed setting [Hz]
R/W
N
High speed setting [Hz]
R/W
N
Maximum rotational frequency
in Hz (can be written only in
Stop condition)
R/W
L
Set Vent Valve status
R/W
L
Valves operating mode
Technical Information
Window Meanings
Admitted Values
0 = AIR
1 = WATER
(default = 0)
0 = NO
1 = YES
(default = 0)
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
19200 = 5
38400 = 6
(default = 4)
Impulse = 0
Continuous = 1
(default = 1)
0 = frequency
1 = power
2 = bearing temp
(default = 0)
350 to WIN 121
(default = 750)
350 to WIN 121
(default = 825)
350 to 825
(default = 825)
1 = open
0 = close
(default = 0)
0 = On Command
1 = Auto Pump Spedd
2 = Autotime
(default = 2)
(A)
17
239/288