Télécharger Imprimer la page

Agilent Technologies Turbo-V 551 Rack Notice De Mode D'emploi page 215

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
Connection J1 Remote I/O Interconnections
Figure 7
All the input/output remote signals to/from the controller must be
connected at P1 mating connector. With the provided J1 mating
connector (shipped with pin 3 shorted with pin 9 and pin 4 shorted
with pin 15) make the connection with AWG 24 (0.25 mm
wire to the pins indicated in the figure to obtain the desired
capability. It is a 15-pins D type connector; the available signals are
detailed in the table, the following paragraphs describe the signal
characteristics and use.
Tab. 3
PIN N.
1
2
3
4
5
6
7
8
9
10
Turbo-V 551 Rack Controller User Manual / 87-901-031-01
Connection J1 Remote I/O Interconnections
Remote I/O connector
SIGNAL
START/STOP (+)
START/STOP (-)
INTERLOCK (+)
INTERLOCK (-)
SPEED SETTING (+)
SPEED SETTING (-)
SOFT START (+)
SOFT START (-)
+24Vdc
SET POINT (RELAY)
Technical Information
INPUT/OUTPUT
IN
IN
IN
IN
IN
IN
IN
IN
OUT
OUT
(A)
17
2
) or smaller
215/288

Publicité

loading

Ce manuel est également adapté pour:

X3501-64001