XD 7000
Proceeding after
a cell has broken
ba77194y01
09/2017
Cause
– Filter test: Instrument defective
– Lamp test:
– If no light leaks out at the rear
side of the photometer:
Lamp defective
– If light does leak out at the rear
side of the photometer
– Wavelength calibration:
– Foreign particle in the cell shaft
– Lens smudged
– Instrument defective
5.2
Actions in the case of a broken cell
WARNING
Cells can contain dangerous substances. If the contents are released,
follow the safety instructions of the package insert. If necessary, take
corresponding protective measures (protective goggles, protective
gloves etc.).
CAUTION
Do not turn the photometer around to poor out the liquid!
When doing so, the liquid could come into contact with electronic
components and damage the photometer.
The photometer has a drain device through which the contents of a broken
cell can drain off without causing any damage.
1 Switch off the photometer and disconnect it from the power supply.
2 Let the liquid drain off into a suitable container and dispose of it properly
according to the instructions of the reagent package.
3 Carefully remove all broken glass, e.g. with tweezers.
Remedy
– Please contact the service
department.
– Replace the lamp
– Carry out a software update
– Remove foreign object
– Clean the lens. If this happens
repeatedly, check the operating
conditions (see Section 8)
– Please contact the service
department.