Manufacturer's EMC guidelines
Electromagnetic immunity
®
The Tel-O-Graph
GSM, Tel-O-Graph
use the Tel-O-Graph
Immunity test
Electrostatic
discharge (ESD) in
accordance with
IEC 61000-4-2
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-11
Magnetic field at
supply frequency
(50/60 Hz) in
accordance with
IEC 61000-4-8
78
®
GSM plus is designed for use in the environment specified below. Only
®
®
GSM, Tel-O-Graph
GSM plus in such an environment.
IEC 60601-1-test level
+/- 6kV contact discharge
+/- 8kV air discharge
3 A/m
Electromagnetic environment -
Compliance level
guidelines
Floors should consist of wood or concrete
+/- 6kV contact
or be tiled with ceramic tiles. If the floor
discharge
consists of synthetic materials, relative
+/- 8kV air discharge
humidity must be at least 30 %.
Not applicable
Not applicable
Not applicable
Magnet fields in network frequency
should correspond with the typical values
3 A/m
found in business and hospital
environments.