5. Putting into use
5.1. Checks
− Check that the measuring instrument is undamaged and that you have received the correct
version.
− Check that the data on the packing slip corresponds with the data shown on the product
received.
If you have any problems and/or in the case of ambiguity:
please contact SIMCO or your regional agent.
5.2. Putting into use
Warning:
− The measuring instrument may be severely damaged when subjected to vibrations or
shocks, or when it is dropped.
− Upon receipt of the measuring instrument the battery is fitted.
− Suspend measurement when the voltage reading is outside the measuring range. If the
range is exceeded, there is a possibility of damaging the sensor.
− It is possible to use this product in ionized air. However, the specified accuracy of within 10%
cannot be guaranteed.
− Do not press on the LCD display.
5.3. Battery check.
Switch the instrument on [POWER]
The battery indicator on the display shows the state of battery.
5.4. Earthing the measuring instrument
The case of the FMX-003 is made of a conductive resin which provides an earth reference for
the measuring circuit. To ensure accurate measurement, the person performing the
measurement must be in good contact with earth, or the instrument must be earthed using the
supplied grounding lead.
FMX - 003 97.52.02.08.20
23
www.DataSheet4U.com
Replace the battery [9.2]
08/08