Courier 5000
NOTE:
• If the scale does not perform the calibration successfully,
previous calibration data is restored.
• After waiting for 40 seconds, if the scale still fails to perform the calibration,
seconds, and the previous calibration data is restored.
4.2.4 Linearity Calibration
The scale performs linearity calibration with 3 calibration points. The full calibration point is established with an
item (Refer to Table 4-1 for the exact weight value of the item) on the platform. The mid calibration point is
established with a weight equal to half of the full calibration weight on the platform. The zero calibration point is
established with no weight on the platform. The full calibration and mid calibration points can be altered during
the calibration procedure.
Calibration Procedures:
Long press the Menu button until you see
1.
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
4.
Press the Yes button, then the display shows
5.
Then the first calibration point and the unit will flash on the display based on the scale capacity and unit
you set in the setup menu. (For example,
skip to step 7.
6.
Change the calibration point:
Short press the No button several times until the desired number appears.
a)
NOTE: press the Back button can decrease the digit.
b)
Short press the Yes button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
Press the Yes button to accept calibration point. It flashes on the display.
d)
7.
Place a calibration mass of the specified weight on the
--C--
shows
8.
The second calibration point and the unit will flash on the display based on the capacity and unit you set in
the setup menu. (For example,
NOTE: After waiting for 40s, if the scale still fails to perform the calibration,
seconds and the previous calibration data is restored.
9.
Steps 6 and 7.
10. The display shows
11. After that the display shows
4.2.5 GEO Code Adjustment
Geographical Adjustment Factor (GEO) code is used to adjust the calibration based on the current location.
Settings from 0 to 31 are available with 12 being the default. Refer to 8.3 Table of GEO Code Values to
determine the GEO factor that corresponds to the terminal's location.
Set the GEO factor:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
4.
The GEO point flashes on the display (For example,
5.
Short press the No button several times until the desired GEO number appears. Press the Yes button to
confirm.
NOTE: Press the Back button can decrease the digit.
6.
Then the display shows
7.
Press the Exit button to exit.
4.2.6 End Cal
ENd
When
is displayed, you can press the Yes button to exit this menu and enter the next sub-menu, or press
the No button to enter to the first menu item in this sub-menu.
Find Quality Products Online at:
.
030.000
--C--
-DonE-
, and then
GEO
. Press the Exit button to exit.
END
.
GlobalTestSupply
www.
CAL E
mM.E.N.U
.
C.A.L
. Press the Yes button.
LIN.
--C--
.
015.000
kg). If you do not need to change the calibration point,
and press the Yes button. The display
platform
kg)
when the Linearity calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
GEO.
12
).
is displayed for 3 seconds and the
CAL E
is displayed for 3
CAL E
is displayed for 3
Press the Yes button.