Lovibond Tintometer Group MD100 Mode D'emploi page 31

Masquer les pouces Voir aussi pour Tintometer Group MD100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 40
Methods
GB
Notes:
1. Filter turbid water samples using filter paper and funnel before analysis.
2. Highly buffered water samples or extreme pH values should be adjusted to a pH of
nearly 7 with 1 mol/l Nitric acid or 1 mol/l Sodium hydroxide.
3. Concentrations above 10 mg/l Cu causes too high test values if the reaction time of 5
minutes is increased. So it is very important to perform the test procedure as described.
4. Substances which may interfere when present in concentrations at:
Aluminium
50 mg/l
Chromium
1000 mg/l
Iron
50 mg/l
Nickel
50 mg/l
Nitrite
all levels
Reagent
Set
VARIO Molybdenum HR1 F10
VARIO Molybdenum HR2 F10
VARIO Molybdenum HR3 F10
MD100_7 07/2016
Form of reagent/Quantity
Powder Pack / 100
Powder Pack / 100
Powder Pack / 100
Order-No.
535300
31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Md100Tintometer md100

Table des Matières