ACE Technik warns: Even if you get a result with the meaning „Safe range", you should not drive
after having consumed alcohol or do any other activities which are badly affected by the
consumption of alcohol.
Usage of personal mouthpiece
If you push the hook aside like shown below, you can use your personal mouthpiece.
Replacing the sensor module
1. Open the battery & sensor cover, to do so raise the cover on the bottom side carefully.
2. Take out the battery.
3. Grab handles and pull down side which indicates on sensor module case.
4. Take out the sensor module.
Don't take up the sensor module when you replace it.
Warning Display
Flow
The breath sample was not strong enough for measurement.
After warming automatically, please blow strong and steady during sampling.
The battery power is low. Please replace the batteries.
bAt
CAL
If you test over 1,000 tests, ACE III will display "CAL" sign for 3 seconds when warming up
for test. This "CAL" sign will be provided the appropriate time for calibration or
replacement sensor module.
Sensor module empty – Please check sensor module.
Err6
System error – Please repair the device.
Err9
Sensor module error – Please replace sensor module.
Err6S
Technical data
Statement of the BAC
Accuracy
Warm-up time
Sensor
Response time
Energy supply
Dimensions of the device
Weight
Ambient conditions
Calibration
All manuals and user guides at all-guides.com
0,00 ~ 4,00 ‰ (fixed basic setting)
+/- 0,005 % BAC (Blood Alcohol Concentration) at 0,05 % BAC
5 seconds up to max. 2 minutes
Fuell cell sensor
Within 10 seconds
2 pieces AAA alkaline batteries (1.5 V)
50 mm x 105 mm x 15 mm
85 g (including batteries)
Operation: 5 - 40° C
Storage: 0 - 50° C
Every 6-12 months or after 500 tests