Only the specified replacement parts should be used for maintenance purposes. The
manufacturer will not be held responsible for any accident occurring following any
repairs made other than by its After Sales Service or approved repairers.
3.1
MAINTENANCE
The clamp must necessarily be away from any conductor and disconnected from the measuring
instrument.
3.1.1
MAGNETIC CIRCUIT
Keep the jaw gap perfectly clean.
Clean the gap with a soft cloth if required.
3.1.2
CASING AND CABLE
Clean the clamp's casing, arms and output cable with a sponge dampened with soapy water.
Rinse these parts with a sponge dampened with clean water.
Never run water over the clamp. Dry with a cloth or pulsed air (at a maximum temperature of
80°C).
3.2
METROLOGICAL VERIFICATION
Like all measuring or testing devices, the instrument must be checked regularly.
This instrument should be checked at least once a year. For checks and calibrations, contact one of
our accredited metrology laboratories (information and contact details available on request), at our
Chauvin Arnoux subsidiary or the branch in your country.
Repairs
For all repairs before or after expiry of warranty, please return the device to your distributor.
3 MAINTENANCE
20