8. Error Messages
Display
Measured temperature too high
Measured temperature too low
* Caution: High fever requires immediate medical attention!
Display / Problem
Display Meaning
Room temperature too high
Room temperature too low
Error function display
Blank display
Dead battery indication
All manuals and user guides at all-guides.com
Meaning
Possible cause and remedy
Displays "H" when measured
temperature is higher than
42.2 °C (108.0 °F). Repeat taking
temperature and confirm with
digital thermometer.
Displays "L" when measured tem-
perature is lower than 32.0 °C
(89.6 °F). See Section 7 for prop-
er technique.
Possible cause and fault
remedy
Displays "H" in conjunction
with the "s" when room temper-
ature is higher than
40.0 °C (104.0 °F).
Display "L" in conjunction
with the "t" when room temper-
ature is lower than
5.0 °C (41.0 °F).
Check battery. Possible damage
from dropping or water.
Please check if battery is
loaded correctly.
If the steady battery icon is the
only symbol shown, the battery
should be replaced immediately.
10
9. Technical Specifications
Type:
Measuring Range:
Accuracy:
Display:
Beeps When:
Memory:
Illuminated Display:
(Back-Lite)
Operating temperature:
Storage/transport temperature:
Automatic Switch-off:
Battery:
Dimensions:
Weight:
Standards:
Please observe the applicable disposal regulations.
Digital Infrared Thermometer IR 1DA1
32.0 - 42.2 °C (89.6 - 108.0 °F)
Laboratory: : ±0.2 °C, 32.0 ~ 42.2 °C
(±0.4 °F, 89.6 ~ 108.0 °F)
Liquid Crystal Display with 0.1 °C (0.1 °F) resolution.
a. The unit is turned ON and ready for the measurement:
1 short "beep" sounds.
b. Measurement complete: 1 long beep sound.
c. System error, malfunction, or "Hi""Low" measurement.
3 short "beep" sounds.
d. Fever alert: If measurement >37.5 °C (>99.5 °F), multiple
beeps sound (instead of a single beep).
Automatically displays the last measured temperature.
a. The display will be lighted for 4 seconds when the unit
is turned ON.
b. The display will be lighted again for 5 seconds when
the measurement has been completed.
-20°C to +50°C (-4°F to +122°F)
-25 °C to +55 °C (-13 °F to 131 °F)
Approx. 1 minute after last measurement has been taken.
2 AAA bateries
122mm (L) x 44 mm (W) x 63 mm (H)
92g (with batteries), 70g (w/o batteries)
Complies with European PrEN12470-5 and U.S. ASTM E-1965
requirements