TECHNICAL FEATURES

Device
Size: 13 X 8 X 2 cm
Weight: 220 g
Container: ABS
Protection level: IP 22
Storage and shipping temperature: from -10°C to 45°C
Max. relative humidity: 30% - 75%
Made in PRC
Charger
Type: FLO Switching Power Adaptor
Model: DK7-065-0200-EU
Input: 100-240 Vac - 50-60 Hz - 0,06 A
Output: 5,8 Vdc - 200 mA
Polarity:
Battery pack: Ni-MH AAA 4.8 V 800 mAh

Use conditions

Temperature: from 0°C to 35°C
Maximum relative humidity: from 15% to 93%
Atmospheric pressure: from 700 hPa to 1060 hPa
The values represent the limits allowed if the product or the accessories
are not in their original package.
Technical features of EMS and TENS currents
Available outputs: Channels 1-2
Constant current: Yes
Intensity: 0-100 mA peak-peak per channel
Waveform: rectangular, biphasic, symmetrical, compensated
Work frequency: 1-150 Hz
Recovery frequency: 1-150 Hz
Pulse amplitude: 50-400 µs
Working time: from 1 to 30 seconds
Recovery time: from 0 to 1 minute
Frequency modulation range: continuous variation from 1 to 150 Hz
Minimum modulation time: 3 seconds
Amplitude modulation range: continuous variation from 50 to 400 µs
Technical features of iontophoresis currents
Available outputs: Channel 1
Constant current: Yes
Min. intensity: 0 mA/1000 Ohm
Maximum intensity: 10 mA/1000 Ohm step 0.1 mA/1000 Ohm
Treatment time: 1 minute
Max. time: 99 minutes