6. Commissioning, operation
Service data (SPDU service protocol data unit)
Service data is always acyclic and exchanged on the request of the IO-Link Master
With the help of the service data, the following parameter values or instrument status can be read:
IO-Link specific:
Index
Object name
(decimal)
16
Vendor name
17
Vendor text
19
Product ID
21
Serial number
24
TAG
33
Last event
40
Process data In
243
Order no
Manufacturer-specific:
Index
Object
(decimal)
name
65
SP1 or FH1
66
RP1 or FL1
WIKA operating instructions pressure switch model PSA-31 for sanitary applications
Format
Access
Visible string
R
Visible string
R
Visible string
R
Visible string
R
max. 16 Byte
R/W
Visible string
Visible string
R
16 Bit unsigned
R
Visible string
R
Format
Access
16 Bit
R/W
unsigned
16 Bit
R/W
unsigned
Factory setting
WIKA Alexander
Wiegand SE & Co. KG
www.wika.com
e.g. 1,013,093
S#
-
-
-
P#
Value range
Factory
setting
1,040 ... 9,000
9,000
1,000 ... 8,960
8,200
Remarks
This SPDU determines the
relationship with the correspond-
ing IODD.
Corresponds to serial number on
the product label (S#).
Customer-specific measuring
point number
Display of the process data that
are cyclically transferred.
Corresponds to product number
on the product label (P#)
Remarks
Switch point/window High
switching output 1
Reset point/window Low
switching output 1
EN