Fault 559
Fault 559
If fault message
when a programme is aborted, the RS232 interface is active and has
to be deactivated.
Open the menu for configuring the RS232 interface and then select
Further settings
/
Interface
Ethernet
Select the
The RS232 interface has now been deactivated. Wait approx.
90 seconds for the module to be operational.
Web interface
The XKM 3000 L Med module has its own web interface, which is
used for accessing general data about the washer-disinfector, chan-
ging network settings and loading updates to the module.
The interface can be accessed from any workstation in your local net-
work. To do so there must be a configured serial interface in the ma-
chine's controls, a current web browser and a network connection to
the machine. The output language is English.
Enter the IP address for your washer-disinfector in the address line
in your browser.
The web interface for the module will continue to be shown in your
browser. Data can take up to 60 seconds to transfer from the washer-
disinfector to the web interface.
The web interface is arranged as follows:
– Appliance information
This page contains general information about the washer-disin-
fector, e.g. model number (Model number), serial number (Serial
number) or MAC Address (MAC address).
– System settings
Data about the interface, such as the IP address (IP address) or the
subnet mask (Subnetmask) can be found on this page. This data
can be changed. It is password protected.
– Password settings
You can change your password on this page.
– Firmware update
If necessary, updates to the module's software can be loaded from
this page. This function is password protected.
The default password is available from the Service department.
If this password has been changed it will revert to the factory default
one if the module is reset.
en - Configuring the interface
appears when the machine is started or
.
option and confirm your choice with OK.
45