METTLER TOLEDO PANTHER/PANTHER PLUS Terminals Installation Guide
Profibus Option
Pluggable Terminal Strip
3-4
(7/03)
The PROFIBUS connection is available at two locations on the PROFIBUS PCB. The first
is a female 9-pin D subminiature connector, which is the PROFIBUS standard
connection. The field connector assembly is not supplied by METTLER TOLEDO. This
connection is the preferred connection in the panel-mount terminal.
Female DE-9
For the PANTHER harsh environment terminal, the pluggable terminal strip must be
used. In these cases, an optional pigtail harness is available to wire from the terminal
strip to a female 9-pin D connector. The adapter harness can be ordered using factory
number 0900-0311. The adapter is shown here.
P1
14852800A Profibus Pigtail Adapter For Desk Mount Jaguar
FROM
TO
PT. 1
END 1
PT. 2
END 2
P1-4
10856700A
---
1/4" STRIP
Terminal strip
9-PIN Connector
1 GND (isolated)
2 N.C.
3 TX/RX+
4 RTS
5 GND (isolated)
6 +5V (isolated)
7 N.C.
8 TX/RX-
9 N.C.
SIGNAL
COLOR
YELLOW
RTS
1 RTS
2 TXD/RXD+
3 TXD/RXD-
4 +5 V (isolated)
5 GND (isolated)
PIN 1