C. CLEANING
1. Fully post processed parts can be cleaned using a dedicated soft brush with
neutral soap and room temperature water.
2. Do not use any abrasive cleaning products on parts printed with BioMed Clear
Resin. Such cleaning products may adversely affect surface finish.
3. After cleaning, always inspect parts for any cracks. Discard if any damage or
cracks are detected.
D. DISINFECTION
Parts may be disinfected in 70% IPA for 5 minutes.
E. STERILIZATION
For sterilization compatibility information, visit support.formlabs.com.
F. STORAGE
1. When not in use, place printed parts in closed, opaque or amber containers.
2. Store in a cool, dry place out of direct sunlight. Excess light exposure overtime
may affect the color of printed parts.
3. Store the cartridges at 10°C - 25°C (50°F - 77°F).
4. Do not exceed 25°C (77°F) when in storage.
5. Keep away from ignition sources.
G. DISPOSAL
1. Any cured resin is non-hazardous and may be disposed of as regular waste.
2. Follow facility protocols for waste that may be considered biohazardous.
a. Liquid resin should be disposed of in accordance with government
regulations (community, regional, national).
b. Contact a licensed professional waste disposal service to dispose of
liquid resin.
c. Do not allow waste to enter storm or sewer drainage systems.
d. Avoid release into the environment.
e. Contaminated packaging: Dispose of as unused product.