air bubbles on the parts. Leave the parts immersed for the amount of time indicated
on the packaging of the disinfectant associated with the concentration chosen for
the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Disinfect the accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5 by
boiling them in water for 10 minutes; use demineralised or distilled water to prevent
calcium deposits.
method C: Disinfect the accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5 with
a hot steam steriliser for baby-bottle (not the microwave type). Perform the process
faithfully following the instructions of the steriliser. To ensure that the disinfection is
effective, choose a steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on
a paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair
dryer).
At the end of each use store the device complete with accessories in a dry place
away from dust.

REPLACING THE FILTER

The device has a suction filter (A3) that should be replaced
when it is dirty or changes color. Do not wash or reuse the
same filter. Regular replacement of the filter is required to
help and ensure correct performance of the compressor.
Check the filter on a regular basis.
Do not replace the filter during use.