Les langues disponibles

5. Commissioning, operation
For information on tapped holes, refer to DIN 3852 or for NPT threads to
ANSI B 1.20.

5.1.1 Tightening torques for compression fittings

Sealing
Stainless steel ferrule
Stainless steel bite-type ferrule
PTFE ferrule

5.2 Electrical connection

Output signal 4 ... 20 mA
Angular connector DIN EN 175301-803
Customer-specific pin assignment see product label
Version with angular connector (DIN EN 175301-803):
For the cable leads we recommend the use of crimped end splices.
To ensure IP65 ingress protection:
Always use the silicon seals
Tighten the locking screw
Ensure the cables are inserted carefully
This is protection class 3 equipment for connection at low voltages,
which are separated from the mains voltage or voltages of greater than
AC 50 V or DC 120 V. Preferably, a connection to an SELV or PELV circuit
is recommended; alternatively protective measures from HD 60346-4-41
(DIN VDE 0100-410).
WIKA operating instructions model TR36
Turns
Max. pressure in bar
1 ¼ ... 1 ½
100
1 ¼ ... 1 ½
100
1 ¼ ... 1 ½
8
With direct sensor output
Angular connector DIN EN 175301-803
EN
15