Maintenance And Cleaning; Cleaning Agent - IKA HB digital 092017 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

- Commands and parameters including successive parameters are
separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data) and each
response are terminated with Blank CR LF (Code: hex 0x20 hex
0x0d hex 0x20 hex 0x0A) and have a maximum length of 80
characters.
- The decimal separator in a number is a dot (Code: hex 0x2E).
The above details correspond as far as possible to the
recommendations of the NAMUR working party (NAMUR
recommendations for the design of electrical plug connections
for analogue and digital signal transmission on individual items of
laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specific IKA commands
serve only as low level commands for communication between the
lab device and the PC. With a suitable terminal or communications
program these commands can be transmitted directly to the
lab device. The IKA software package, labworldsoft
a convenient tool for controlling lab device and collecting data
under MS Windows, and includes graphical entry features, for
motor speed ramps for example.
The following table summarises the (NAMUR) commands
understood by the IKA equipment.
The device is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.
Cleaning
Only use cleaning materials recommended by IKA:
Dirt

Cleaning agent

Dyes
Isopropyl alcohol
Building materials Water containing detergent/isopropyl alcohol
Cosmetics
Water containing detergent/isopropyl alcohol
Food
Water containing detergent
Fuels
Water containing detergent
Other materials
Please consult IKA
Wear protective gloves during cleaning the devices.
Electrical devices may not be placed in the cleansing agent for the
purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for clean-
ing or decontamination, you must ascertain with IKA that this
method does not destroy the device.
, provides
®

Maintenance and cleaning

Remove the device from the
mains before cleaning.
NAMUR Commands Function
IN_NAME
Read the device name
IN_PV_2
Read medium temperature actual value
IN_SP_2
Read medium temperature set value
OUT_SP_2 V
Set medium temperature set value
IN_SP_3
Read safety temperature set value
OUT_SP_3 V
Set safety temperature value
IN_SP_74
Read medium type set value
(1 - water, 0 - oil)
OUT_SP_74 V
Set medium type value
(1 - water, 0 - oil)
START_2
Start heating
STOP_2
Stop heating
USB cable A - B
This cable is used to connect the USB interface to a PC.
USB Micro B
Ordering spare parts
When ordering spare parts, please give:
- Device type,
- Serial number, see rating plate,
- Position number and description of spare part, see
www.ika.com,
- Software version.
Repairs
Please only send devices in for repair that have been cleaned
and are free of materials which might present health haz-
ards.
For this, use the "certificate of compliance" form which you
can obtain from IKA or can download a version for printing from
the IKA website at www.ika.com.
If your appliance requires repair, return it in its original packaging.
Storage packaging is not sufficient when sending the device - also
use appropriate transport packaging.
USB A
9

Publicité

Table des Matières
loading

Table des Matières