Doseur de Fluide Haute Précision Ultimus V
Annexe B — Protocole RS-232 (suite)
2.6.3.10 Real Time Clock Read Command
This command returns the current time of the real time clock on the dispenser.
Client command and data: (STX) + 04 + EE-- + D2 + (ETX)
The above text packet in hexadecimal format:
STX
0
0x02
0x30
NOTE: "D0" is the letter "D" followed by the digit "0."
Return Format: D0HhhMmmAMa
hh:
Hours. 0–23 for 24 hour format, 1–12 for 12 hour format
mm: Minutes. 0–59
a:
Hour format. 0 = AM, 1 = PM, 2 = 24 hour format
EXAMPLE: If the current time on the dispenser is 14:25, the text packet received from the dispenser is:
(STX) + 0B + D0H14M25AM2 + F9 + (ETX)
The above text packet in hexadecimal format:
STX
0
0x02
0x30
F
9
0x46
0x39
70