Cleaning and
Disinfection:
Automated
Cleaning:
Manual
Disinfection:
Manual (Inter-
mediate-Level)
Packaging
Sterilization
Drying
Maintenance
Storage
Manufacturer
Contact
Do not use automated washer/disinfectors for reprocessing
SmartLite
Focus
®
®
damage will occur.
1. The SmartLite
Focus
®
to be cleaned by scrubbing with a pre-impregnated wipe or dis-
posable towel soaked with an alcohol-based, tuberculocidal,
quaternary ammonium solution with a label claim for cleaning
(e.g., VoloWipes disinfecting/cleaning/deodorizing Wipes) or
pH-neutral, phosphate-free cleaning solution (e.g., Dr. Schu-
macher Instru Plus [3%]) according to local regulations and use
according to cleaning solution manufacturer's Instructions for
Use until it is free of visible residues. Remove all visible soil,
ensuring fluid penetrates all crevices. Do not allow solution to
penetrate the casing. Discard used wipes. Additional wipes may
be used.
2. Allow the device to air dry.
1. After cleaning wipe all device surfaces with a new single use
cloth in combination with an alcohol-based, tuberculocidal, qua-
ternary ammonium solution e.g.: VoloWipes disinfecting / clean-
ing/deodorizing Wipes, BBraun Meliseptol
according to local regulations, and use according to disinfectant
solution manufacturer's Instructions for Use. Pay special atten-
tion to seams, areas around buttons, window, and crevices.
2. Allow the device to air dry.
3. Wipe the devices with a sterile, clean, lint-free cloth dampened
with distilled water.
4. Pay special attention to device seams and insertions.
No particular requirements.
Sterilization is not allowed. No methods have been validated.
Do not subject components to Steam autoclaving or liquid chemi-
cal sterilant immersion. Component damage will occur.
Wipe the devices dry with a sterile, clean, lint-free cloth. Allow the
components to fully air dry before storage.
Visually inspect to ensure that all contamination has been re-
moved.
Visually inspect power supply and cord for damage.
Components that are damaged, worn, or distorted should be dis-
carded.
No additional maintenance or lubrication is recommended.
Store the SmartLite
at room temperature, away from moisture or excessive humidity.
Within the United States, call Dentsply Sirona at 1-302-422-4511.
For areas outside the United States, contact your local Dentsply
Sirona representative.
light handpiece or charging base. Component
light handpiece and charging base have
®
Focus
light handpiece and charging base
®
®
Rapid, approved
®
9