Zero Adjusment (Rel) - Chauvin Arnoux C.A 6630 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
English
select the resistance and voltage ranges using the V-RANGE and
Ω Ω Ω Ω -RANGE keys.
read the measurement results (resistance and DC voltage) directly on the
display unit.
Remarks :
if the reading is Outside the Limits for the range (resistance and
-
voltage), "OL" lights.
in case of malfunction of the AC test current, "----" is displayed.
-
4.3

ZERO ADJUSMENT (REL)

The zero adjustment (or measurement lead compensation) function serves to
quantify the influence of the measurement leads and deduct it from the
measurement results.
The results of the zero adjustment will then be taken as the measurement zero
(origin).
- short-circuit the red and black
measurement contact tips as shown
opposite
- press the REL key: R is then displayed and
the resistance and voltage values are set to
zero
- the connections to the battery to be tested
can now be carried out.
Warning :
The lead compensation applies only to the
measurement range selected beforehand
and only until the instrument is switched off.
31