Storage
The sensor can be stored for several months filled with
O
electrolyte (order no. 30 298 424) as long as the
2
protective cap is in place. In order to avoid the 6-hour
polarization time before reuse of the sensor, it can be
kept stored connected to the METTLER TOLEDO O
Sensor-Master InPro 6800.
Attention! The electrolyte should be replaced before
use if the storage period exceeds 3 months.
If the storage period exceeds 6 months, the sensor
should be stored dry, i.e. without any electrolyte in the
membrane body.
Attention! A sensor being stored dry (without
electrolyte in the membrane body) may on no account
be connected to the O
2
8
Product specification
8.1
Certificates
Each sensor is delivered with a set of 3.1 B certificates
(complying with EN 10204.3/1.B).
All wetted metal parts (sensor shaft, cap sleeve and
membrane body) are identified with a engraved
symbol corresponding to the heat number on the
paper certificate delivered with the sensor.
Each wetted metal part (sensor shaft, cap sleeve and
membrane body) is polished in order to get a surface
roughness lower than 0.4 µm (16 µin). This represents
a roughness grade number of N5 (according to ISO
1320:1992).