Command
asel
atest
Calibration and adjustment
cdate
crh
crhclr
ct
ctclr
ctext
fcrh
l
li
Other commands
filt
frestore
reset
Description
Show or set analog output parameters and scaling.
Test analog outputs by forcing them to a given value.
Show or set adjustment date.
User calibration for humidity measurement.
Clear user calibration for humidity measurement.
Factory calibration remains.
User calibration for T measurement.
Clear user calibration for temperature measurement. Factory
calibration remains.
Show or set adjustment information text.
Two-point calibration after humidity sensor change.
Show adjustment offset and gain.
Set adjustment offset and gain.
Show or set measurement filtering.
Restore factory settings. Clears all user settings, factory calibration
remains.
Reset the device.