Autoclave Procedure:
1. Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth. Do not use a wire brush.
2. Lubricate with PANA SPRAY Plus. Refer to 5. Lubrication.
3. Insert into an autoclave pouch. Seal the pouch.
4. Autoclavable up to a max. 135˚C.
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5. Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121˚C for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
CAUTION
• Do not heat or cool the handpiece quickly. Rapid change in temperature could break the glass rod give abnormal strain to other metals.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.