The accuracy of an electronic measuring instrument is more constant if the operating temperature remains the same.
For this reason, these PE balances are built so that when the instrument is switched off using the single control bar (1), only the display is switched off. The balance
electronic components remain switched on as long as the power cable is connected (stand-by operation).
It is also important that the balance display always be switched on (by pressing the control bar) with the pan empty.
Switching the balance on
– Whit the pan empty, briefly press the single
control bar (1); all display segments light up
for several seconds:
This allows you to make a check of the display.
Then the display indicates zero, with as many
decimal places as the balance model readability
permits, e.g.:
0.000 g with the PE360 or
0.00 g with the PE3600.
6
Switching off the display
– Lift single control bar (1); the display is
switched off.
OFF Display
If there is a power interruption at any time, "OFF" will
appear in the display as soon as power is restored.