4. Design and function / 5. Special conditions ...
Through the high protection class of the thermometer (IP65) and its
liquid damping, operation under vibration conditions is possible.
EN

4.2 Scope of delivery

Cross-check scope of delivery with delivery note.

5. Special conditions for use (X conditions)

1) Design temperatures
Ambient temperature: -20 ... +60 °C
Medium temperature: max. +600 °C
(instruments with liquid filling: max. +250 °C)
Observe the surface temperature for ATEX application:
The permissible medium temperature does not only depend on
the instrument design, but also on the ignition temperature of the
surrounding gases, vapours or dust. Take both aspects into account.
2) Maximum surface temperature
For instruments with liquid filling, make sure that the medium
temperature is lower than 250 °C.
The surface temperature mainly depends on the medium
temperature. For prevention, consider the maximum medium
temperature as maximum surface temperature.
10
WIKA operating instructions models 53, 54, 55 (ATEX)