Insert the finger sensor connector into the pulse oximeter (when the sensor is not inserted, the display
views the following writing: Prb). It is sufficient to make the grooves present on the connector and on
the socket match.
Instrument description and symbols
Select
Pushbutton for the selection of SpO 2 Max and Min values and PULSE/min.
Pushbutton to increase SpO 2 Max and Min values and PULSE/min.
Pushbutton to decrease SpO 2 Max and Min values and PULSE/min.
Turn on/off audio signals
Indicator for disabled audio signals.
Run-down battery indicator. When the battery charge is low, the battery indicator lights up
and gives out two beeps. In this case it is necessary to replace the batteries. At this point the
run-down battery indicator turns off.
Prb
Sensor not connected with the equipment
Pot
No patient connected
Lo
Low signal
Operating instructions
To turn the equipment on, press the green ON/OFF pushbutton. To confirm power-up, an acoustic
signal is released and PWR pilot light turned on. The right display views the following writing: Pot
(no patient connected).
Insert the finger of the patient into the examination sensor. It is possible to use any finger, however it
is advisable to select the finger suitable to the sensor dimensions. (The PULSE OXIMETER can be
fitted with different sensors, follow the instructions).
In order to perform an accurate detection of the value, the patient shall keep still.
Wait 6 seconds and the sensor starts up measuring. The left display shows the value of blood oxygen
saturation rate (SpO 2 %) and the right display shows the heart rate (Pulse/min).
COMDEK PULSE OXIMETER is provided with a sound signal warning that the values detected by
the sensor are higher or lower than the ones stored in the equipment.
Default alarm settings are:
SpO 2
Max
SpO 2
Min
Heart rate
Max
Heart rate
Min
OPERATION
100%
50%
250 bpm
30 bpm