Fig. 51
Inserting
lamp with
cover
a right
b wrong
a
b
• Insert the new 12 V 100 W lamp (Fig. 51) with
the dust cover straight into the socket un-
til it stops. Be sure that the lamp is inserted
straight.
• Remove the lamp's dust cover.
Fig. 52
Rear view, Leica DMI4000 B and DMI6000 B
1
Installation point for lamp housing mount
or mirror housing
2
Holes for lamp housing mount or mirror housing screws
2
2
1
• Replace the housing and fasten it in place us-
ing the fastening screw.
6.11 Installing the Lamp Housing Mount and
Mirror Housing (Leica DMI4000 B and
DMI6000 B)
Place lamp housing mount (Fig. 53) or mirror
housing on rear wall. Mount from front with sock-
et head screws.
Fig. 53
Lamp housing mount
Fig. 54
Lamp housing 106z
1
Collector adjustment
2
Vertical lamp adjustment
3
Horizontal lamp adjustment
4
Adapter ring
4
6. Assembly
2
3
1
47