Mettler Toledo InPro Serie Instructions D'utilisation page 25

Masquer les pouces Voir aussi pour InPro Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

InPro 6850i / 6900i / 6950i O
Note: make sure that all air bubbles are removed
from the membrane body. Air bubbles can be
removed by carefully tapping on the membrane
body.
10. Slip the membrane body over the interior body
while holding the sensor in a vertical position.
The excess electrolyte will be displaced and have
to be removed with a paper tissue.
Important! No electrolyte, sample media or
contamination may be present between the
membrane body and the cap sleeve. Please
check carefully!
11. Carefully slip the cap sleeve over the fitted
membrane body, holding the sensor in a vertical
position and screw it tight. The cap sleeve must
be clean and dry.
12. After each exchange of electrolyte or membrane
body, the sensor has to be repolarized and
recalibrated.
© 02 / 09 Mettler-Toledo AG
Printed in Switzerland
Sensor 12 / 25 mm
2
1
2
1
O
Electrolyte
2
2
InPro 6850i, 6900i, 6950i
25
52 206 349

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Inpro 6850iInpro 6900iInpro 6950i

Table des Matières