4.5.3 Calibration Settings

Press the Settings Button to access the settings mode.
Icon
Description
Cal Curve
EN
Standard Set
Reminder
Default Cal
Post Verify
Cal Checker
not TB350 
1) 
26
4 Operation
Option
T-CAL
®
User
Formazin
Default
Unique
On
1)
Off
 
On
Off
On
(Default)
Off
Function
Selects prepared stabilized formazine
standard at pre-defined concentrations
Selects user prepared formazine standard
at pre-defined concentrations
Used if no unique identifier is set.
Unique identifier of standard set.
Reminds user that calibration is due.
Option to force Calibration before
additional measurements occur.
Default setting.
Option to return to factory calibration
setting.
Requires user to perform a Verification
after calibration is complete.
Default setting.
During the calibration procedure, if an
individual standard is out of tolerance, the
user must correct the standard to proceed
or cancel the calibration procedure.
Feature prevents user from a calibration
error due to an incorrect standard. The log
screen will have a line item for Calibration
Checker and it will show an "On" or
"Enabled" state.
During the calibration procedure, if an
individual standard is out of tolerance,
user will not be notified that a calibrator
is out of Tintometers predefined range.
The log screen will have a line item for
Calibration Checker and it will show an
"Off" or "Disabled" state.
EN   TB 350