Technical data – Operating messages
GB
Light source:
Battery:
Auto-OFF:
Ambient conditions: 5–40°C
CE:
To ensure maximum accuracy of test results, always use the reagent systems
supplied by the instrument manufacturer .
Lo
E 01
E 10 / E 11
E 20 /E 21
E 22
E 70
E 71
E 72
E 73
E 74
E 75
32

Technical Data

LED, Filter ( = 528 nm)
9 V-block battery (Life approx . 600 tests),
without display light .
Automatic switch off 5 minutes after last keypress
30–90% rel . humidity (non-condensing)
Certificate for Declaration of CE-Conformity
at www .tintometer .de

Operating messages

Measuring range exceeded or excessive turbidity .
Result below the lowest limit of the measuring range .
Replace 9 V battery, no further tests possible .

Error codes

Light absorption too great . Reasons: e .g . dirty optics .
Calibration factor "out of range"
Too much light reaching the detector .
Battery capacity was too low during measurement . Change battery .
CLt:
Factory calibration incorrect / erase
CLt:
User calibration incorrect / erase
CLP:
Factory calibration incorrect / erase
CLP:
User calibration incorrect / erase
pH:
Factory calibration incorrect / erase
pH:
User calibration incorrect / erase
MiniDirect_ClpH_1 09/2008