Warnings
The UV lamp and the display screen contain mercury and may be considered hazardous
material in your local area. Dispose of these items in accordance with federal, state, or local
government regulations.
Should the UV lamp become broken or damaged it should be handled in accordance with
your organization's toxic waste handling procedure and disposed of in accordance with
federal, state, or local government regulations.
To protect against accidental exposure to ultra-violet radiation, do not operate the UV lamp
outside of its protective housing. The UV lamp housing may be hot; use extreme caution
whenever touching the UV lamp housing.
Always stop analysis before turning off the Analyzer.
This is a Class A product. In a domestic environment, this product may cause electromagnetic
interference in which case the user may be required to take adequate measures to correct the
interference.
To avoid false TOC readings and possible damage to the Analyzer, always make sure the
sample inlet is open and the DI water cartridge is filled before starting analysis. Always close
the door of the Standard iOS System when not sampling from a vial.
Make sure the DI water cartridge is full, particularly when running samples with high TOC or
high salt concentrations. Always "clean-up" the Analyzer by running low-TOC DI water after
running high TOC or salt samples.
If this instrument is used in a manner not specified by SUEZ, the protection provided by the
instrument may be impaired.
The glass window in the UV reactor indicates proper lamp operation and should not be viewed
too closely (closer than 6 inches) or for extended periods (greater than 1 minute). Protective
eyewear is recommended.