4.2.1
Cleaning and disinfecting procedure
1. Remove the respiratory filters.
2. Clean all parts with lukewarm water adding Sekusept
and using a soft cloth (max. temperature: 86 °F/30 °C, concentra-
tion according to degree of soiling: 0.5 - 1 %)
3. Thoroughly rinse all parts under running water.
4. Prepare a disinfectant bath containing water and Incidin
(max. temperature:86 °F/30 °C, concentration: 1.5 %)
5. Place all parts to be disinfected into the disinfectant bath (duration:
15 minutes).
6. Thoroughly rinse all parts under running water.
7. Dry all parts in the air or in a drying cabinet (max. temperature:
140 °F/60 °C). Do not expose to direct sunlight.
®
1)
Sekusept
is a registered trademark of Ecolab Deutschland GmbH
®
2)
Incidin
is a registered trademark of Ecolab USA Inc.
X-plore 3000
Maintenance
®
Cleaner
1)
.
®
Rapid
2)
.
27