Technical Specifications; Standard Delivery; Delivery Versions; Error List - Renfert Mobiloskop S Manuel D'utilisation

Masquer les pouces Voir aussi pour Mobiloskop S:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 7
6.
Spare Parts
You can find components subject to wear and the spare parts on the spare part list in the internet at
EN
www.renfert.com/p918.
The components excluded from the warranty (such as consumables or parts subject to wear and tear)
are marked on the spare part list.
Serial number and date of manufacturing are shown on the type plate of the unit.
7.
Guarantee
Provided the unit is properly used, Renfert warrants all components of the Mobiloskop S for a period
of 3 years. The original sales receipt from the specialist dealer must be presented with any warranty
claims. Components subject to natural wear are excluded from this warranty.
The warranty is voided in case of improper use; failure to observe the operating, cleaning, mainte-
nance, and connection instructions; in case of independent repairs or repairs by unauthorized person-
nel; if spare parts from other manufacturers are employed, or; in case of unusual influences or influ-
ences not in compliance with the utilization instructions. Warranty service shall not extend the original
warranty.
8.

Technical Specifications

Light path:
Visual field diameter:
Focal depth:
Working distance, approx.:
Weight, approx.:
9.

Standard Delivery

1 Mobiloskop S
1 Support arm
1 Lens retainer
1 Lens with eyepieces
1 Protecting glass
1 Cover
1 Instruction manual

10. Delivery Versions

No. 2200-0110
No. 2200-0200
No. 3-0743
No. 2200-5000
No. 2200-0420

11. Error List

Problem
No visible image.
Image is not sharply
focussed.
The support arm does
not stand still.
44 mm [1.734 inch] (5 x), 22 mm [0.867 inch] (10 x)
22 mm [0.867 inch] (5 x), 10 mm [0.394 inch] (10 x)
Eyepiece pair, 10 x
Additional 2 x lens, magnifies from 5 x /10 x to 10 x / 20 x
Replacement protecting glass to protect the lenses
Mobiloscope lighting
Stand with screw clamp
Possible cause
• Magnification change knob
incorrectly adjusted.
• Incorrect dioptre compensation
setting.
• Incorrect distance to the object.
• Dirty lens / protecting glass.
• Fixation is loose.
- 2 -
45°, angled
150 mm [5.9 inch]
1.8 kg [4.0 lbs]
Corrective action
• Turn the knob to its end stop.
• Repeat the dioptre compensation procedure.
• Change the distance to the object.
• Use a clean, lint free cloth to clean the lens /
protecting glass.
• Tighten fixation.
Subject to modification.