Produktgruppe 512
47 von 48
Use a beaker or measuring cup to refill the container to the fill line.
Determine the dispensed amount in the beaker/cup and enter it in the display (V:.....ml).
Press the diskette symbol to save the dispensed amount.
Register the refill message (e.g., dispensing pump 1).
Operator level Main menu Settings Dispenser refill message Dispensing pump 1: No
one level) Diskette symbol to save the refill message.
Check dispensing in all programs and adjust as necessary.
Register water recycling (WRG)
Service Settings Basic settings Water intake Valve block 2 Water valve 3 : K6 OK Back 2x (to go back 2 levels)
Diskette symbol to save WRG registration.
Carry out a functional check.
Back up all data via service mode.
Service Backup Backup data for EW Overwrite existing data? Diskette symbol to overwrite EW data.
To quit service mode, switch off the machine.
Finish up:
Check water connections for leaks.
Re-install the liquid dispensing cover on the rear panel.
Re-install the machine lid.
Carry out a safety check.
Reconnect the machine to the electrical and gas supplies and switch it on.
am
10.02.2012
Diese Unterlagen dürfen ohne unsere Genehmigung weder vervielfältigt noch Dritten zugänglich gemacht werden. Eigentumsrechte vorbehalten.
Umbau- und Montageanweisung
M.-Nr. 09081940
Yes OK Back (to go back