Troubleshooting Guide; Technical Support And Repairs 12 Speciications; Classiications; Disposal Of Unit - DENTSPLY Cavitron Select SPS Instructions D'utilisation

Masquer les pouces Voir aussi pour Cavitron Select SPS:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Section 12: Troubleshooting
Although service and repair of the Cavitron
®
Select™ SPS™ should
be performed by DENTSPLY
personnel, the following are some
®
basic trouble shooting procedures that will help avoid unnecessary
service calls. Generally, check all lines and connections to and from
the System, a loose plug or connection will often create problems.
Check the settings on the System's knobs.

12.1 Troubleshooting Guide

System will not operate:
(Power Indicator Light is not lighted.)
1. Check that the Power Control knob is in the ON position,
and that the detachable Power Cord is fully seated in the
receptacle on back of the Cavitron
Select™ SPS™ System.
®
2. Check that the System's three-prong plug is fully seated in an
appropriate AC receptacle, and that AC current is present.
3. Check that the DC power plug is properly inserted into the
DC input located on the back of the System.
(Power Indicator Light is lighted.)
1. Check that the Foot Control Connector is fully seated in the
Foot Control Receptacle on the back of the System.
2. Check that the Handpiece Cable Assembly is fully seated in
the Handpiece Receptacle.
ENGLISH • 16
System operates:
(No lavage low at insert tip.)
1. Assure that Lavage control is properly adjusted.
2. Check that irrigant supply control valve/s (dental oice
water supply) are open.
3. Check that the Handpiece Cable Assembly is fully seated
in the Handpiece Receptacle.
4. Check to see if Water Line or Pump Lavage Filter is
discolored—if so, replace.
5. Check to assure that liquid is available in reservoir bottle.
12.2 Technical Support and Repairs
For technical support and assistance call 1-800-989-8826 Monday
through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other
areas, contact your local DENTSPLY
®
representative.
Section 13: Speciications
Cavitron
Select™: 100-240 VAC Systems
®
Electrical Voltage 30 VDC +/- 5%
Current
2 amps DC Max
Frequency
50/60 Hertz
Water Pressure
20 – 40 psig
Flow Rate
Minimum setting (CCW) 0 – 15 ml/min
Maximum setting (CW) >45 ml/min
Weight
425 g / 15 oz.
Dimensions
Height:
2.4 in.
Width:
3-1/8 in.
Depth:
7 1/2 in.
Handpiece Cable:
72 in.
Foot Control Cable:
72 in.
Power Cord:
2 m
Water Supply Line:
96 in.
Cavitron
®
Select™ Power Supply (674022001):
Electrical Voltage 100 – 240 VAC
Current
1,6 amps Max @ 100 – 240 VAC
Insulation Requirement
Reinforce
Frequency
47 – 63 Hertz
Weight
550 g / 19.4 oz. Max
Dimensions
Height:
14,3 cm / 5.7 in.
Width:
7,6 cm / 3.0 in.
Depth:
14,3 cm / 5.7 in.
Power Cord:
213 cm / 84 in.
Footswitch
Not for operating theatres
Protection Class IPX1
Cavitron
®
Select™ Reservoir Pump:
Electrical Voltage 30 VDC +/- 5%
Pump Pressure
15 – 19 psig
Flow Rate
40 ml/min
Weight
513 g / 18 oz.
Dimensions
Height:
10,2 cm / 4.0 in.
Width:
8,4 cm / 3.3 in.
Length:
30 cm / 11.8 in.
Operating Environment:
Temperature:
15 to 40 Deg. Celsius
Relative Humidity:
30% to 75% (non-condensing)
Transport and Storage Conditions:
Temperature:
0 to 70 Deg. Celsius
Relative Humidity:
10% to 95% (non-condensing)
Atmospheric Pressure:
500 to 1060 hPa
Section 14: Classiications
• Type of protection against electric shock:
• Degree of protection against electric shock:
• Degree of protection against the harmful ingress of water:
• Mode of operation:
• Degree of safety of application in the presence of a lammable Equipment not suitable for use in the presence
anaesthetic mixture with air or with oxygen or nitrous oxide: of lammable anesthetics or oxygen.
• According to medical device directive:
Section 15: Disposal of Unit
• Accordance with local and state laws.
Symbology
DIRECT CURRENT
TYPE B EQUIPMENT
FOOTSWITCH
CAUTION, CONSULT
ACCOMPANYING DOCUMENTS
CONSULT INSTRUCTIONS FOR
USE
This symbol is a mandatory marking for
devices entering the European market to
indicate conformity with the essential health
and safety requirements set out in European
Directives. The symbol may be accompanied
by a four-digit identiication number of the
notiied body.
Dispose of in accordance with the Waste
Electrical and Electronic Equipment Directive
2002/96/EC of the European Parliament and
the Council of the European Union
WEEE
C
US
EUR
USA
Class 1
Type B
Ordinary
Continuous
IIA (rule 9)
ENGLISH • 17

Publicité

Table des Matières
loading

Table des Matières