• You can establish the connection as illus-
trated at the front sockets or with the plug
clamps on the back (19/20).
• Activate all DC outputs with the button
„OUTPUT ON".
• All output voltages and currents are con-
trolled via the „master" device.
Sensor operation „sense"
The sensor operation „sense" allows the precise voltage setting directly on the consumer. This reliably
compensates a possible voltage drop via the connection cables. Sensor operation is only possible in
normal (master) mode.
• Switch the power unit off.
• Connect the consumer with the output sockets (observe the cor-
rect polarity!) or the plug clamps (O/P) on the power unit.
• Connect the sensor cable from the connection clamps of the con-
sumer with the sensor input on the power supply observing the
right polarity. The positive lead must be connected to the clamp
„RS+" (18) and the negative lead to the clamp „RS -" (21).
To do so, press the little slot
inwards with a screwdriver and
insert the insulated conductor into
the round opening.
• Switch on the mains power supply at the power switch (1). The
voltage and current display appears after a short initiation period.
• • Set the parameters according to your specifications as
described in the chapter „Commissioning".
• Check once again that the output voltage was set correctly and
activate the DC output with the button „OUTPUT ON".
• Now you can switch on the connected consumer.
2