Specifications; General Specifications - Chauvin Arnoux C. A 6410 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1

GENERAL SPECIFICATIONS

Conformity to the EN 6110-1 Ed 95 standard:
Instrument completely protected by double insulation
EN 61010-2-032 Ed 95, 150 V, category I I I and degree of pollution 2 (class 2).
Leakage line and distance in the air: 3.3 mm minimum
Emission: NF EN 55081-1 Ed 92
Immunity: up to 10V/m, level 3 as per NF EN 55082-1 Ed 97
Limiting overloads:
Case: polycarbonate
Dimensions: 55 x 100 x 240 mm
Max. Clamping diameter: 32 mm Ø
Weight: approx. 1 kg
Watertightness: IP30, Group III equipment as per EN 60529 Ed 92
Power supply: - 9 V alkaline battery 6LF22 or equivalent
- average consumption: approx. 40 mA
- average service life: approx. 8 hours,
Note: extreme environmental conditions may disturb the internal microprocessor.
Simply disconnecting the battery may be sufficient to eradicate this malfunction.

SPECIFICATIONS

permanent current 100 A max. (50/60 Hz)
transient currents (< 5 s) 200 A (50/60 Hz)
IK04, as per EN 50102 Ed 95
i.e. 1000 measurements of 30 seconds.
With a rechageable NiCad battery, the average service life is
approx. 400 measurements of 30 seconds.
42

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C. a 6412C. a 6415C. a 6415r

Table des Matières