OPERATION

ON/OFF OPERATION
ON/OFF switches ON, OFF, and gives access to the secondary functions of the
instrument.
When switched ON, if the ON button is still pressed, the whole display is lit. After being
pressed for 2 seconds, the remaining service life of the battery is displayed as a %
(display of flashing
symbol). Model C.A 6415 also states the number of
MEM
recordings that have already been carried out (displays flashing
symbol).
As soon as it is switched ON, the clamp configures itself for resistance measurement (W ).
The buzzer function is activated or not (on models C.A 6412, C.A 6415 or C.A 6415R),
depending on its state when the clamp was last turned off. This is the same for the
memory and alarm functions on models C.A 6415 and C.A 6415R.

CHOICE OF THE MEASUREMENT UNIT

Measurement in W W W W W
After pressing ON or after switching on, the instrument performs a resistance
measurement.
OL
on the main display indicates that the value of resistance measured exceeds the
measurement range.
----- on the main display indicates either that the clamp is open, or that dirt is stopping
the clamp from closing correctly.The
symbol is present on the display.
Measurement in A (models C.A 6412, C.A 6415 and C.A 6415R only)
After A is pressed, the instrument performs a current measurement.
OL
on the main display indicates that the value of the current measured exceeds the
measurement range.

HOLD DIGITAL VALUE ON THE DISPLAY

This function is accessible with the HOLD button, only on measurement mode.
Press HOLD to freeze the last measurement displayed. The
HOLD
symbol is
displayed and the A, W and AL (models C.A 6412, C.A 6415 and C.A 6415R only)
buttons are inoperative. Recording of the measurement is possible if memory mode is
ON (C.A 6415 and C.A 6415R exclusively). In this case, the NOISE and R < .1W W W W W symbol
are also recorded in memory if displayed.
Press the HOLD button again to reset the instrument to instantaneous measurement
mode and
HOLD
disappears from the display.