Factory default settings

15.1 Fixed settings for relays, LED and
alarm
Setting
Fault relay
Energized / Normally Open
Yellow Fault LED
Lit if a fault message is available
Red Alarm LED
Single blink if A1 condition is present.
Double blink if A2 condition is
present. If an alarm is configured as
acknowledgeable and the alarm is
acknowledged, single/double blink
changes into continuously lit.
Alarm Hierarchy
A2 overrides A1 on the LED. However,
the A1 and A2 relays operate inde-
pendently. I.e. if A1 is acknowledgeable
and A2 is not, and the gas concentra-
tion is such that it triggers A1 and A2:
acknowledging will cause the A1 relay
to release. However, the red LED will
still double blink as long as the A2 con-
dition continues to exist.
15.2 Settings which can be changed via
the menu
®
15.2.1 Polytron
8xx0
Menu
Default setting
Relay active on
Active on alarm
alarm / No alarm
A1 alarm: direction Rising (falling for
O
Sensors)
2
A2 alarm: direction Rising
A1 latch mode
Non-latching
A2 latch mode
Latching
A1 acknowledg-
Acknowledgeable
ment mode
A2 acknowledg-
Not acknowledge-
ment mode
able
Normal operation
Energized
A1 relay
Normal operation
A2 relay
Calibration pass-
_ _ _ 1
word
Settings password _ _ _ 2
LCD setting
On
SIL status
Off
Language
EN
Instructions for use
|
Dräger Polytron
Range
On / Off
Falling / Rising
Latching / Non-
latching
Acknowledgeable /
Not acknowledge-
able / Pre-acknowl-
edgeable
Energized / Not
energized
On / Off
On / Off
DE / EN / FR / ES /
RU / ZH
®
8000 Series
Menu
Default setting
Auto-calibration
Off
Function key
Faults
Interface
Default setting
Fault current
1.2 mA
Warning
Off
Warning current
3.0 mA
Warning cycle
10 sec
interval T1
Warning cycle
1 sec
interval T2 (warn-
ing current)
Maintenance sig-
static
nal
Maintenance cur-
3.4 mA
rent
Analog offset
0 mA
Analog span
0 mA
0
®
HART
address
126
1)
Modbus address
1)
even (default)
Modbus parity
Modbus bau-
9600 bps
1)
drate
®
126
PROFIBUS
1)
address
1) Fieldbus interfaces are not in accordance with the BVS 13 ATEX
G001 X. Using the fieldbus interface no relays are available.

15.3 Sensor-specific values

See sensor data sheet for further specifications.
®
15.3.1 Polytron
8100 EC
Menu
Default setting
A1 Alarm
Depending on the
sensor
A2 Alarm
Depending on the
sensor
Hysteresis for A1
0
alarm at direction
rising
Hysteresis for A2
0
alarm at direction
rising
|
Factory default settings
Range
On / Off
Graph, fault, warn-
ing, vitality, bump
test
Range
0 to 3.5 mA
On / Off
0 to 3.5 mA
5 to 60 sec
1 to (T1-1) sec
static / dynamic
0 to 3.5 mA
-0.5 to 0.5 mA
-0.5 to 0.5 mA
0 to 15
2-126
even, odd parity, no
parity
9600 or 19200 bps
2-126
Range
0 to A1
0 to A2
en
31