Cleaning, Disinfection and Sterilization
Operation
2
Pre-Disinfection
3
Rinsing
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Operating mode
•
Pulp and dentine
remnants must be
removed from the
accessories
immediately.
•
After using the
accessories on the
patient, place them
directly into a dish
filled with a suitable
cleaning and
disinfection solution
(e.g. CIDEZYME
®
ENZOL
Enzymatic
Detergent Solutions,
Johnson & Johnson
Medical, 0.8% for
between 1 minute
and 2 hours) for
cleaning, pre-
disinfection and
interim storage.
•
Wash the
accessories under
flowing sterile,
deionized water or in
a disinfection solution
at least three times
for a minute each
time, in order to
remove all visible
traces of
contamination and
remnants.
•
Abundant rinsing
(at least 1 min).
•
Do not allow remnants to dry
on. Clean no later than 2
hours. For visible impurities
that are observed on
instruments, a pre- cleaning
is recommended by brushing
them manually with soft
material. Only use clean, soft
brushes to manually remove
contamination and remnants,
or a clean, soft cloth or wipe
that is only used for this
purpose. Do not use metal
®
,
brushes or wire wool. Check
that no visible contamination
or remnants remain, and
repeat the pre-cleaning
process if necessary.
•
Make sure that the products
are fully immersed.
•
The disinfectant solution
should be aldehyde free (to
avoid blood impurities
fixation), suitable for
disinfecting the accessories,
and compatible with the
accessories.
•
Disinfectant used for pre-
treatment is only for personal
protection and does not
obviate the need for
disinfection once cleaning
has been completed.
Pre-treatment should never
be omitted.
•
Use clean water.
•
If a pre-disinfectant solution
contains a corrosion
inhibitor, it is recommended
to rinse the instruments just
before the cleaning.
Warning
25/222