IMPORTANT NOTICE
WHEN YOU HAVE EATEN, DRUNK ALCOHOL OR, SMOKED A
CIGARETTE, WAIT AT LEAST 15 MINUTES BEFORE STARTING
THE MEASUREMENT. ANY TEST CONDUCTED PRIOR TO THE
EXPIRATION OF THE RECOMMENDED WAITING PERIOD WILL
RESULT IN INACCURATE RESULTS.
IF THE TEST RESULT IS BELOW 0.10‰ (PER MILL = g/L) OR
0.01% BAC (US setting) OR 0.05mg/L (UK settting) THEN THE
DEVICE SHOWS 0.00 (‰ OR % BAC OR mg/L – depending on
the setting) ON THE DISPLAY. THIS HAPPENS BECAUSE AT A
LOW RANGE OF ALCOHOL CONCENTRATION IT IS POSSIBLE
TO MEASURE OTHER MATERIALS IN YOUR MOUTH.
This breathalyzer is a high quality and sensitive measurement
device. To maintain the accuracy of the sensor, it should regu-
larly be used. If the device is not used for an extended period,
the sensor can dry out. ACE Instruments recommends using the
breathalyzer at least 2 times per month (even if you have not
been drinking alcohol).
Due to changes in the product, the procedures stated can
change. This can take place without prior notice or adaptation
of the operating manual and is no cause for complaint, return or
price reduction. The user is responsible for recognizing discre-
pancies between the operating manual and the actual handling
of the breathalyzer and must react accordingly.
Only Use Original ACE Instruments Mouthpieces
Guarantees test result accuracy
Š
Guarantees hygiene of you and the device
Š
Guarantees full warranty of this device
Š
ACE Instruments advises
even if you get a result meaning „safe range", you should
never drive after consuming alcohol or do any other activities
that are severely impaired by alcohol consumption and might
endanger your health and safety or the health and safety of
others.
21
Display Indications
Display
Meaning
The battery is empty.
BAt
The testing cannot
be performed
because the battery
is low.
FLo
The sampling
volume or blow
power is not
enough to analyze
the breath sample.
Out
If no breath sample
was given within
3 minutes, the
device turns off
automatically.
Outc
The testing cannot
be performed
because the device
is turned on outside
of the operating
temperature.
Err2
"Err2" states that
fluid or saliva has
entered the device
and the sensor is
unable to evaluate it.
500
The total number
of breath tests per-
formed has reached
500. The device
should be calibrated
promptly to maintain
the warranty and
accuracy.
Action
Replace with new
batteries.
Replace with new
batteries.
Press the start button, the
countdown will start again,
and "BLOW" will appear
on the display. You can
now perform a new test.
Press power button to
restart the countdown
cycle and retest.
Repeat the test after
the unit has reached
a temperature within
the operation range
(5°C to 40°C).
Wait 20 minutes and
then try again. If "Err2" is
displayed for a long time,
contact your seller for
customer service.
Contact your seller for a
calibration service. You
can send the device for
calibration to the service
address specified at
www.alkomat.net.