Pull the filter cap [5] forwards and turn it.
1
Note:
If the filter remains in the unit when the cap is re-
moved, remove the filter from the unit using twee-
zers or similar.
Replace the filter cap [5] containing the new filter.
2
Check that the filter cap is firmly in place.
3
8

Cleaning and sterilisation

Atomiser and accessories
Warning
Adhere to the following hygiene instruc-
tions to avoid possible damage to health.
The nebuliser and accessories are in-
tended for multiple use. Please note
that cleaning requirements differ ac-
cording to the different application ar-
eas:
Notes:
The atomiser and its accessories must
not be mechanically cleaned with brus-
hes or similar, otherwise irreparable
damage may occur and the success of
the therapy can no longer be guaran-
teed.
For additional requirements regarding
the necessary hygienic preparations
(hand washing, handling of medicines
or inhalation solutions) in high-risk
groups (e.g. cystic fibrosis patients),
contact your doctor.
Ensure that all parts are sufficiently
dried each time after cleaning and ster-
ilisation. Residual moisture or water
droplets can result in an increased risk
of bacterial growth.

Preparation

Directly following each treatment, all
parts of the atomiser and any acces-
sories used must be cleaned of medi-
cine residues and contaminants.
To do this, dismantle the nebuliser into its
component parts.
12
11
13
14
26
Remove the mouthpiece [10]
from the nebuliser.
Remove the medicine con-
tainer [14] from the upper part
of the nebuliser.
Remove the nozzle attach-
ment [13] from the medicine
container [14].
Open the sealing cap [12] on
the atomiser.
The nebuliser is later reas-
sembled in the reverse or-
der.