Operation
4.3.5 Acid Density
Electrolyte density
In the
the density of the battery's electrolyte, which
is used for the initial filling (standard value
for non-tropical countries 1.28 kg/l).
The standard value 1.28 kg/l (standard set-
ting ex works) is neither shown nor printed
out. The value is shown for settings between
1.22 kg/l and 1.27 kg/l, and this also appears
in the printout.
i
The electrolyte density setting will exert an
influence on the battery test result.

4.3.6 Date

Using the battery tester software, you can
change the time in the
menu. Adjust the time using the 
and the 
key
ged by a maximum of ±12 hours. The date
changes automatically when the time is chan-
ged to below 0.00 or more than 24.00 hours.
Save the new time with
i
If the "Date" menu is not shown when in con-
figuration mode, or if the date and time are
not printed out on the results protocol, the
complete clock module must be exchanged.
1 689 979 868 (2005-10-07)
menu you can set
Configuration/Date
key
(down). The time can be chan-
E.
Enter
4.4
Last Test Result
The last Battery-Test executed can be retrie-
ved in the menu item
last test result remains saved in the memory
even after disconnecting the tester from the
battery.
(up)
0 684 400 703

Last Test Result.